September 9, 2008
ULTRAHIGH FREQUENCY SOUND©
Lyn Paul Taylor, A.A., B.A., M.A., R.P.T.
(Editing Assistant and Computer Consultant: Joanna Soon, B.S.)
Ultrahigh
frequency sound (ultrasound) is a form of energy derived from
mechanical vibration. It is produced when a crystal set in an ultrasonic
transducer is electrically stimulated causing the electrical energy to be
changed into a mechanical vibration that produces a focused beam of ultrasonic
waves. The waveform can be continuous or pulsed
(intermittent). The ultrasonic waves pass through a metal cap that covers the
crystal (together called the sound head) and, under the right conditions, are
passed into biological tissues.
The depth of penetration of
ultrasound into the tissues depends on the frequency of the sound (the number
of vibrations per second) and the density of the tissues. The
therapeutic effects of ultrasound are said to depend on frequency,
intensity (amplitude), and duration of application.
There are two basic types of ultrasound heads,
moveable and stationary. The moveable ultrasound head was developed before the
arrival of the pulsed waveform. Because of the heating effect
produced by the continuous waveform, comfort dictated that the ultrasound head
be continuously moved over a fairly large area (several times the ultrasound
head size) to reduce thermal effects on the tissues sounded (the sound head
heated up). The pulsed (intermittently interrupted) ultrasound
waveform does not have the same tissue heating effect and can sometimes be
applied with a stationary (fixed) ultrasound head, or with relatively little
sound head movement. This allows the ultrasound to be concentrated
in a more concise area. Sound heads vary in size, ranging from four inches to
less than one half inch in diameter.
A coupling agent is necessary to conduct the
ultrasound from the sound head to the skin because of the minuscule air gap
that always exists between the uncoupled sound head and the skin, even when the
sound head is firmly pressed against the skin. The coupling agent's job is
essentially to seal the sound head and the skin together. The ideal coupling
agent should be fluid enough to allow for free transmission of the sound into
the skin but viscous enough to prevent run off if it warms up during
application. Substances as fluid as mineral oil, or as viscous as soluble
colloid gels, have been shown to be adequate couplants.
It is axiomatic that the longer ultrasound passes
through a tissue, the greater the thermal and nonthermal
physiological effects. Application duration of ultrasound in clinical practice
usually ranges from four to eight minutes, varying as the size of the treatment
site varies. As a rule of thumb, an area varying from 16 to 72 cm2
is sounded for six minutes and more than 72 cm2
for eight minutes. Smaller areas may be sounded for two to four minutes.

An ultrahigh
frequency sound instrument
featuring 1 Mhz
and 3 Mhz heads
The lower the frequency of ultrasound, the deeper
the penetration of significant levels of sound into biological tissues.
Absorption (attenuation) of the ultrasonic beam decreases as the frequency
decreases. At three megahertz (Mhz), only 50% of the ultrasonic beam reaches a soft tissue depth of one centimeter (cm), and only 3% reaches
a depth of five cm. At one Mhz (the frequency most commonly used
by ultrasound units produced in the United States), 82% of the ultrasonic beam
penetrates to a soft tissue depth of one cm and 32% reaches five cm. Lower
frequencies, of course, penetrate deeper, but the depth of most soft tissues
(until bone is reached) would seem to make greater penetration generally
unnecessary and probably undesirable since greater penetration increases the
risk of stimulating uninvolved structures (especially if treating abdominal
structures).
Intensity (current
amplitude) is the amount of power (watts or W) applied to a definitive surface
area (per cm2). Intensities applied therapeutically range from 0.6
to 2 W/cm2. As the intensity of the ultrasound increases,
the temperature of the tissues being sounded rises in proportion to the amount
of sound absorbed by the respective tissue; the greater the tissue absorption
of the sound, the greater the heating. Early researchers found that lower
treatment intensities often produced better recipient responses than higher
levels. This follows the Amd-Schulz Law that states that low dosages of any form of energy are likely to produce beneficial physiological
reactions within stimulated tissues, while medium to high dosages (more than
2.0 W/cm2) may produce pathological results (heightened
sensitivity or pain). Levels below 0.6 W/cm2
have little or no therapeutic value.
The density of the tissues
will affect attenuation of the ultrasound beam, with greater densities causing more attenuation then less dense tissues. The degree of attenuation varies
from one type of tissues to another. Blood attenuates the ultrasound beam at a
rate of 3% per cm, fat 13%, muscle 24%, blood vessel 32%, skin 39%, tendon 59%,
cartilage 68% and bone 96%.
The number of treatments within a given time span
plays an important role in the effectiveness of ultrasound treatment. Some
authorities have recommended that ultrasound treatment of a given treatment
site be conducted as frequently as twice a day (as close as one half hour
apart). As a rule, the more frequently a site can be treated, the greater the
effect of the ultrasound. It should be noted, however, that different
conditions will require a different frequency of treatment, and the total
number of treatments required for successful resolution will also vary.
Ultrasound can be applied in a continuous
or pulsed waveform. The continuous waveform has the greater thermal effect and has been reported to be effective
at increasing collagen tissue elasticity, altering blood flow, changing nerve
conduction velocity, increasing pain threshold and enzyme activity, and
changing contractile activity in skeletal muscles. The pulsed waveform
has reportedly been used when the nonthermal mechanisms of ultrasound can be of
benefit and when heating needs are minimized (as in the treatment of stasis
ulcers). The nonthermal mechanisms include acoustical streaming, increasing cellular permeability, exertion of a propelling force to drive
chemicals across the cell wall barriers, and the dispersion of fluids. The
thermal effect of the pulsed waveform is minimized because when in the off-phase
of the pulse, the intensity is zero. Any expression of intensity, when
speaking of the pulsed waveform, must be an average of the temporal peak value
of the intensity and the zero value of the off-phase.
Application:
- The timer, waveform and dosage levels of the ultrasound machine
should be preset.
- The coupling agent should be liberally applied over the treatment
site.
- The sound head should be placed on the coupling agent covering the treatment site with the flat surface of the sound head against the
skin (in bony prominent areas this may only be possible with a small sound
head).
- The ultrasound unit should be turned on.
- If a moveable sound head and a continuous waveform are
being used, the sound head should be moved over the treatment site (being
careful to keep the sound head flat against the skin) with a continuous
circuitous motion, or with continuous stroking back and forth motion. The sound
head should be moved at a slow, steady rate. If a stationary sound head and a
pulsed waveform are used, no motion of the sound head is necessary, but the
practitioner should remain in constant attendance and the amplitude adjusted to
maintain recipient comfort.
- Following treatment, the sound head should be thoroughly cleansed
in preparation for the next application.

Applying a coupling agent in preparation for Ultrasound treatment

Ultrasounding a treatment site on the low back
Precautions:
Because blood supply to the
lens is poor, therapeutic levels of ultrasound should not be
applied over the eye. The poor circulation doesn't allow the
heat generated by ultrasound to dissipate adequately and may lead to cataract
formation. The retina may also sustain damage from therapeutic levels of
ultrasound because any ultrasound entering the eye will be only slightly
attenuated by the aqueous and vitreous humors, allowing the ultrasound to have
full impact on delicate retinal tissues.
Therapeutic
levels of ultrasound should not be applied over a pregnant
uterus. Animal studies have suggested that the temperature elevation
produced by an ultrasound beam may cause low birth weight, brain size reduction, and various orthopedic deformities.
Ultrasound applied over the testes may
produce a prolonged elevation of testicular temperature that may result in temporary
sterility.
Malignant tissue should not be ultrasounded. In vitro studies have suggested the
possibility that ultrasound may promote malignant cellular detachment and
metastasis.
When treating an immature
recipient, caution should be exerted to avoid exposing long bone ends to
ultrasound because of the remote possibility that the epiphysis
may be damaged by exposure to ultrasound. It should be noted, however, that
the literature reviewed suggested that epiphyseal lines are, in fact, safe from normal therapeutic levels of ultrasound. The evidence suggests that only
ultrasound intensities above 3.0 W/cm2,
applied with a stationary sound head for periods of three minutes
or more may damage epiphyseal plates and retard bone growth (also true for
demineralization of the bone).
Thrombus formations (blood clots) should
not be ultrasounded because of the danger of increasing thrombus formation and promoting emboli production.
Caution should be exercised
when applying ultrasound to areas that suffer from inadequate circulation. Elevated tissue temperatures may threaten such tissues since the heat transfer
normally provided by the circulatory system is missing. In such cases, the continuous
waveform of ultrasound should be avoided and only
the pulsed waveform used and that with caution.
Ultrasound should not
be applied over areas containing prosthetic implants or other hardware. Any stimulated metal implant may expand from the heat
imparted to it by ultrasound. This may cause it to push out against the
supporting tissues (bone), only to shrink again when the heat has dissipated,
leaving the implant loose within supportive tissues. As a consequence, the
implant may lose its stability and require removal and replacement (if
possible).
References:
W.T. Coakley, "Biophysical Effects of Ultrasound
at Therapeutic Intensities", Physiotherapy, 64:6, June 1978. Pp.
166-168
M. Dyson, J.B. Pond, J. Joseph, and R. Warwick,
"Stimulation of Tissue Regeneration by Pulsed Plane-Wave Ultrasound",
IEEE Transactions on Sonics and Ultrasonics, July 1970. Pp. 133-139
M.
Dyson and J. Suckling, "Stimulation of Tissue repair by Ultrasound: a
Survey of the Mechanisms Involved", Physiotherapy, 64:4, 1978. Pp.
105-108
C.S. Enwemeka, "The Effects of Therapeutic
Ultrasound on Tendon Healing", Am J Phys Med Rehabil, vol. 68,
1989. Pp. 283-287
B.H. Ferguson, A Practitioners Guide to the
Ultrasonic Therapy Equipment Standard, U.S. Department of Health and Human
Services, Public Health Service, Food and Drug Administration, Rockville,
Maryland, July 1985.
G.T. Haar, "Basic Physics of Therapeutic
Ultrasound", Physiotherapy, 64:4, April 1978. Pp. 100-104
A. Hartley, Therapeutic Ultrasound, Anne Hartley
Agency, Etobiocoke, Ontario, 1987.
J.F. Lehmann, B.J. DeLateur and D.R. Silverman,
"Selective Heating effects of Ultrasound in human beings", Archives
of Physical Medicine and Rehabilitation, June 1966.
F.E. Miller and J.B. Weaver, "Ultrasound
Therapy", The Physical Therapy Review, 34:11, 1954. p. 562
E.M. Oakley, "Application of Continuous Beam
Ultrasound at Therapeutic Levels", Physiotherapy, 64:6, June 1978.
Pp. 169-172
D.C. Reid and G.E. Cummings, "Efficiency of
Ultrasound Coupling Agents", Physiotherapy, 63:8, August 1977. Pp.
255-257
H.P. Schwan and E.L. Carstensen, "Advantages and
Limitations of Ultrasonics in Medicine", JAMA, vol. 149, May 1952.
Pp. 121-125
L.P. Taylor, T. Hui, The Taylor Technique of Soft
Tissue Management, Inflammation: Evaluation & Treatment, 2002. Pp.
44-63
Therapeutic Ultrasound, Reprinted from Physiotherapy,
March/April, 1987, Journal of the Chartered Society of Physiotherapy, London,
England.
F.S. Zach, B. Boynton, K. Phillips, and E. Smith,
"Localized Application of Ultrasonic Energy", British Journal of
Physical Medicine, August 1957.
M.C. Ziskin and S.L.
Michlovitz, "Therapeutic Ultrasound", Thermal Agents in
Rehabilitation, F.A. Davis Co., Philadelphia, Pa., 1986. Pp. 141
[Back to Articles]